Original Contributions: Rapid Estimation of Carboxyhemoglobin Level in Fire Fighters

Richard D. Stewart, MD; R. Scot Stewart; William Stamm; Richard P.Seelen

Of the occupational groups at risk of exposure to carbon monoxide, fire fighters are subjected repeatedly to the highest, most hazardous concentrations of the gas. Smoke can contain CO in concentrations of 0.1% to 10%, and a few deep breaths of it can elevate the blood carboxyhemoglobin (COHb) to levels that produce a severe hypoxic stress on the cardiovascular system (Table).1-4 The stress of this CO-induced hypoxia is further intensified when it is superimposed on the stress of maximum physical exertion, heat stress, pulmonary insult from a myriad of toxic inhalants, and possible underlying coronary heart disease. The complexity of the matter is confounded further when a fire fighter is re-exposed to CO on multiple occasions during a 24-hour tour of duty, permitting the accumulation of an excessive, occasionally lethal, quantity of COHb.5, 6

Efforts to develop a suitable on-the-job method that will detect a toxic amount of COHb before a hazardous concentration has been reached have not been successful to date, to our knowledge.6, 7 To prevent injury from CO and other smoke constituents, heavy reliance has been placed on the routine use of respirators. Nevertheless, over 10,000 fire fighters are overcome by toxic gases each year.5

What is needed is a practical field method to determine COHb accurately and rapidly. Ideally, the analytical equipment should be portable so that it could be used at the scene of a fire. The method should also be simple enough so that it could be performed by a non-chemist technician.

This communication reports the results of a study in which fire fighters were trained to use a rapid electrochemical technique to measure the CO in their expired breath and then estimate the magnitude of their exposure to the gas.

Experimental Method

One hundred seventy fire fighters of the First Battalion, Milwaukee Fire Department, volunteered to participate in this study. Two Emergency Medical Technicians from each shift were given six hours of instruction in the collection of breath samples, sample analysis, and the interpretation of breath CO data.

1

To collect the breath samples, the subjects were instructed to hold their breath for twenty seconds, then discard the first portion of their expired breath and collect the last portion (alveolar) in a 5-liter Saran bag (Figure 1).8-10 The breath samples were immediately analyzed in a properly calibrated breath analyzer (model 2100 Ecolyzer), a 4-kg portable instrument whose operating principle is based on the electrochemical oxidation of CO at a Teflon-bonded diffusion electrode.11 2 The elapsed time for breath sample collection and analysis averaged 1 1/2 minutes. To establish the breath CO-blood COHb relationship presented in Figure 2, 71 paired alveolar breath and venous blood samples were collected from 56 fire fighters for CO and COHb analysis. The theoretical relationship between the amount of CO in the expired breath and the COHb in the blood shown in Figure 2 was derived as follows: The Haldane equation expresses the relationship between COHb, CO tension, oxyhemoglobin, oxygen tension, and the affinity of CO for hemoglobin (M) in a blood sample at equilibrium.

1

During the 20 seconds of breath holding, the tension of CO in the alveoli approximates the CO tension in the pulmonary capillaries. The oxygen tension stabilizes at approximately 80 mm Hg in Milwaukee. With this information, the Haldane equation can be solved for alveolar CO concentration in parts per million where COb is the alveolar CO in parts per million, COHb is the percentage of carboxyhemoglobin saturation; O2Hb is the percentage of oxyhemoglobin; PB is the barometric pressure in millimeters of mercury. In the calculations, M was assigned a value of 210; PCO2 was assigned a value of 80.

The calibration of the breath analyzer was checked each day by the emergency medical technicians, using a compressed air-CO mixture. This calibration was re-checked five times a week by a technician from The Medical College of Wisconsin.

Results

2a

The emergency medical technicians were able to analyze breath samples for CO, using the breath analyzer after the three training sessions. The COHb levels after the first fire fighting episode of a 24-hour tour of duty were as follows: In 55 non-smoking fire fighters, the mean COGb level was 5.0% (range, 1.4% to 9.1%). The five highest COHb values in this group were 7.9%, 8.4%, 8.6%, 8.6%, and 9.1%. In the remaining 56 cigarette smokers, the mean COHb level was 7.0% (range, 2.9% to 13.0%). The five highest COHb values were 10.4%, 10.7%, 11.8%, 12.4%, and 13.0%. The non-smoking firement had COHb saturations comparable to those of heavy cigarette smokers, while five of the cigarette smokers had COHb elevations above 10% saturation.

As an accuracy check on the breath analyzer method, the CO in a random number of breath samples was determined by gas chromatography.12 The average difference in the results between the two methods was 1.2 ppm over the range of 1.4 to 132 ppm, confirming the accuracy of this electrochemical method. Using the CO-COHb relationship shown in Figure 2, the standard error of estimating COHb, given a breath CO concentration of 8 or 50 ppm, was +/- 0.5% COHb saturation. The fire fighters enthusiastically endorsed this method for estimating their COHb body burden. They quickly learned the proper breath collection procedure and did not resent spending the few minutes required to perform the COHb screening test.

There are compelling reasons why fire fighters who are exposed to potentially toxic concentrations of CO each day should be screened for the presence of elevated blood COHb each time exposure occurs. With any increase in COHb and the resultant decrease in blood oxygen-carrying capacity, there is a compensatory increase in cardiac output and a diversion of blood to those organs most sensitive to oxygen lack.13 Diseases hearts in subjects required to exert maximally may not possess the reserve to increase cardiac output to desired levels without ischemia.3, 4 Therefore, a means to detect toxic amounts of absorbed CO is required if unnecessary cardiac stress is to be prevented.

1a

To be of greatest value, any COHb detection method employed should be available for use immediately following each exposure to CO. This permits the prompt initiation of oxygen therapy in cases of overexposure. In addition, the method should be rapid, accurate, economical, and noninvasive. The described electrochemical breath analysis performed by the firemen themselves meets these criteria. The portability of the instruments and its self-contained power supply for eight hours of operation make possible at-the-site analyses with the frequency necessary to define the cumulative CO exposure.

While the CO-COHb relationship shown in Figure 2 should be applicable to the majority of geographical locations, a more precise CO-COHb correlation curve can be constructed using the equation given. Both barometric pressure and altitude, with their effect on the partial pressure of oxygen in the pulmonary capillaries, are important determinants.

From the standpoint of having firemen perform their own analyses, the only disadvantage of this electrochemical method is the requirement that periodic calibration checks be made. Fortunately, the linearity of the sensor response necessitates only a one-point calibration procedure.11 In field use, this can be achieved easily each day, using a commercial compressed air-CO standard.

Sjostrand14 was the first to demonstrate that COHb saturation could be estimated from alveolar CO concentration. The relationship between COHb and the magnitude of exposure was carefully set forth by Coburn, who utilized breath CO as a measure of endogenous CO production.15 Prior to this study, the best equation for the COHb-CO relationship was that presented by Peterson and Stewart.16 However, the expired breath method failed to gain in popularity because a practical field method could not be devised. Efforts to monitor individual CO exposures continued to rely on rapid blood COHb methods, none of which were ideal for use in the field.12, 17, 18 The development of a portable electrochemical cell to make Sjostrand’s CO breath measurement on a sample collected, using the 20-second breath-holding technique of Jones et al, 8 should now provide that long sought-after biological test of CO exposure for all occupational groups.

The current monitoring of environmental exposure for American industry set forth in the NIOSH Criteria and Recommended Standard for Occupational Exposure to CO recommends breathing zone measurements to define average and peak exposure concentrations “at least annually” and COHb determinations only when needed to evaluate “borderline exposure.”19 With the technology available in 1972, this was an acceptable surveillance program. However, all recognized its salient weakness: an absolute inability to detect CO overexposure in occupational groups such as fire fighters in time to effectively combat hypoxic stress. The use of the described rapid electrochemical method of breath analysis remedies this defect in the surveillance program. The quantity of CO absorbed by any workman can be determined by breath analysis performed with a frequency necessary to define the magnitude of the CO exposure.

This investigation was supported in part by The Johnson Wax Fund. Jack E. Peterson, MD, suggested the use of the Ecolyser breath analyzer, and Energetics Science, Inc, Elmsford, New York, provided the Ecolyser; and the Fire Fighters of the First Batallion, Milwaukee Fire Department, volunteered to evaluate this screening test.

References

1. Stewart RD, Peterson JE, Fisher TN, et al: Experimental human exposure to high concentrations of carbon monoxide. Arch Environ Health 26: 1-7, 1973.
2. Anderson RF, Allensworth DC, DeGrott WJ: Myocardial toxicity from carbon monoxide poisoning. Ann Intern Med 67: 1172-1182, 1967.
3. Aronow WS, Isbell MW: Carbon Monoxide effect on exercise-induced angina pectoris. Ann Intern Med 79: 392-395, 1973.
4. Anderson EW, Andelman RJ, Strauch JM, et al: Effect of low-level carbon monoxide exposure on onset and duration of angina pectoris. Ann Intern Med 79: 46-50, 1973.
5. 1973 annual death and injury survey. Internation Fire Fighter 59: 6-9, 1974.
6. Wise MK. Denver draws blood to check carbon monoxide. Fire Engineering 119: 28-31, 1966.
7. Sammons JH, Coleman RL: Firefighters’ occupational exposure to carbon monoxide. J Occup Med 16: 543-546, 1974.
8. Jones RH, Ellicott MF, Cadigan JB, et al: The relationship between blood carbon monoxide concentrations during breath-holding: simple estimate of COHb saturation. J Lab Clin Med 51: 553-564, 1958.
9. Ringold A, et al: Estimating recent carbon monoxide exposures: a rapid method. Arch Environ Health 5: 308-318, 1962.
10. Stewart RD: The use of breath analysis in clinical toxicology, in Hayes W (ed): Essays in Toxicology. New York, Academic Press, 1974, vol. 5, chap. 5.
11. Bay HW, Blurton KF, Sedlak JM, et al: Electrochemical technique for the measurement of carbon monoxide. Anal Chem 46: 1837-1839, 1974.
12. Collison HA, Rodkey FL, O’Neal JD: Determination of carbon monoxide in blood by gas chromatography. Clin Chem 14: 162-171, 1968.
13. Paulson OB, Parving H, Olesen J, et al: Influence of carbon monoxide and of hemodilution on cerebral blood flow and blood gases in man. J Apple Physiol 35: 111-116, 1973.
14. Sjostrand T: A method for the determination of carboxyhemoglobin concentrations by analysis of alveolar air. Acta Physical Scand 16: 201-207, 1948.
15. Coburn RF, Forster RE, Kane PB: Considerations of the physiological variables that determine blood carboxyhemoglobin concentrations in man. J Clin Invest 44: 1899-1910, 1965.
16. Peterson JE, Stewart RD: Absorption and elimination of carbon monoxide by inactive young men. Arch Environ Health 21: 165-1711 1970.
17. Henderson M, Apthorp GH: Rapid method for estimation of carbon monoxide in blood. Br Med J 2: 1853-1854, 1960.
18. Dahms TE, Horvath SM: Rapid, accurate technique for determination of carbon monoxide in blood. Clin Chem 20: 533-537, 1974.
19. Criteria and recommended standard for Occupational Exposure to Carbon Monoxide, report no. HSM 73-11000. Cincinnati, National Institute for Occupational Safety and Health, 1972.